How do I reboot the EPC in my mass spectrometer? - WKB7776
Article number: 7776
ENVIRONMENT
- Xevo TQ-MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-XS
- Xevo TQ-S micro
- ACQUITY TQD
- ACQUITY SQD
- ACQUITY QDa
- 3100 Mass Detector
- Quattro Micro
- Quattro Premier
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
- Synapt G2
- Synapt G2-S
- Synapt G2-Si
- Synapt XS
ANSWER
First use Telnet to restart the EPC. All mass spectrometers have a reset button for the EPC, but if not done correctly, the system may vent.
Resetting the EPC using Telnet:
- Click Windows.
- In the search bar, type "Telnet epc" and press Enter.
- If Telnet is currently configured on your computer, the DOS prompt window launches.
- The window should contain a ">", followed by a blinking cursor.
- Type "reboot" in lowercase and press Enter.
- The EPC is now rebooting, so you can close the Telnet window. Depending on the MS, the reboot can take up to five minutes. A chime should signal completion.
ADDITIONAL INFORMATION
id7776, A-SQ, A-SQ, MALDIG2SI, SQD, SQD2, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD