Poor mass accuracy on SYNAPT G2-Si - WKB17015
Article number: 17015
SYMPTOMS
- Performed calibration just before the runs.
- Mass accuracy is off even by 20 ppm or more.
ENVIRONMENT
SYNAPT G2-Si
CAUSE
Resolution on low e4 ions is below the set point. You can find the set point under System > Acquisition settings.
FIX or WORKAROUND
- Tune up the resolution on low e4 ions as described in this article.
- Ensure that the lock mass signal is at least 5e5 counts per second in continuum mode.
ADDITIONAL INFORMATION
id17015, MALDIG2SI, SYNG2SIHD