Troubleshooting BioAccord instrument setup failures (Guided Troubleshooting) - WKB95208
Article number: 95208
SYMPTOMS
- Instrument setup won't run
- Instrument setup fails
- Calibration fails
- Manually infusing lockmass solution yields weak or no signal
- "Low Response Detected" error message
ENVIRONMENT
- BioAccord System
- ACQUITY RDa Detector
STRATEGY
- Make sure to use the correct LockSpray solution - WKB119274
- If setup won't run, ensure that your vacuum level is <7e-7 mbar - WKB72092
- Ensure that fluidics solutions are not empty and that the fluidics tubing is fully submerged in the fluidics solutions - WKB53359
- Purge the syringe a few times - WKB53253
- If some calibration peaks are missing, sonicate the calibration solution - WKB72090
- Depending on which step failed:
ADDITIONAL INFORMATION