Instrument setup for RDa fails - WKB53253

Article number: 53253

SYMPTOMS

  • In UNIFI, instrument setup fails for RDa
  • No beam is seen in UNIFI

ENVIRONMENT

  • RDa detector
  • UNIFI workstation

CAUSE

Air in the fluidics lines

FIX or WORKAROUND

  1. Perform additional purges of the syringe.
  2. Perform instrument setup once the air is removed from the lines and a beam is seen.
  3. Ensure peek tubing is submerged in calibrant

ADDITIONAL INFORMATION

 