BioAccord System acquisition stops in UNIFI with "lockmass check" error - WKB53355

Article number: 53355

SYMPTOMS

  • Acquisition stops in UNIFI with "lockmass check" error
  • Lockmass intensity is seen at about 1e7

ENVIRONMENT

  • UNIFI Workstation
  • RDa detector
  • BioAccord System

CAUSE

Lockmass signal is too strong.

FIX or WORKAROUND

  1. Ensure that lockmass concentration is appropriately made in the 2e6 range.
  2. Restart sample acquisition after confirming appropriate lockmass intensity.

ADDITIONAL INFORMATION

 

id53355, RDA