RDa instrument setup or calibration fails in UNIFI - WKB53359
Article number: 53359
SYMPTOMS
- RDa calibration fails in UNIFI
ENVIRONMENT
- RDa detector
- UNIFI Workstation
CAUSE
The fluidics tubing is above the level of the solution in the bottle, there is air in lines due to a loose fitting or the piston pump is failing
FIX or WORKAROUND
- Reinstall the bottle so that the fluidics tubing is submerged in the lockmass or calibration solution
- Make sure all fluidics fittings are tight at the divert valve
- If there is still a delay in calibration solution infusing (30s or greater) consider replacing the piston pump (700012108)
- Rerun the instrument setup/calibration