RDa calibration fails in all modes - WKB53365

Article number: 53365

SYMPTOMS

  • RDa calibration fails in both positive and negative ion modes
  • Results show several missing peaks

ENVIRONMENT

  • ACQUITY RDa Detector
  • UNIFI Workstation

CAUSE

Bad calibration solution

FIX or WORKAROUND

  1. Put on a new bottle of calibration solution (part number 186009013.)
  2. Rerun the calibration.

ADDITIONAL INFORMATION