"Blocked Aperture" error in UNIFI for RDa - WKB53240
Article number: 53240
SYMPTOMS
- Error in UNIFI: "Blocked Aperture"
- Backing pressure readback is lower than expected at 1.0 mbar
ENVIRONMENT
- RDa detector
CAUSE
Blocked aperture
FIX or WORKAROUND
- Sonicate the aperture.
- Reinstall the aperture and confirm good backing pressure (usually 1.4 mbar).
- If sonication can't restore the normal backing vacuum (resolve the issue), replace Aperture Disc (P/N 700012106)
Refer to Section 3.8 in the Overview and Maintenance Guide
ADDITIONAL INFORMATION
id53240, RDA