BioAccord repeatedly fails instrument setup - WKB119274

Article number: 119274

SYMPTOMS

  • Instrument setup fails
  • New instrument - so it is clean
  • Repeated attempts of instrument setup also fail

ENVIRONMENT

  • BioAccord
  • ACQUITY RDa
  • waters_connect/UNIFI 1.9.9 and later
  • RDA 2.0 driver or later

CAUSE

The wrong LockSpray solution is being used.

FIX or WORKAROUND

Use the waters_connect LockSpray solution (186009298).

ADDITIONAL INFORMATION

 

id119274, RDA