Skip to main content
Waters

RDa setup fails or does not run - WKB72092

Article number: 72092

SYMPTOMS

  • Setup fails or does not run
  • Vacuum OK but was recently vented

ENVIRONMENT

  • RDa
  • BioAccord

CAUSE

Vacuum allows the RDa to enter Operate but is not yet low enough for the software to run the setup successfully.

FIX or WORKAROUND

Wait at least an hour and try again.

ADDITIONAL INFORMATION

On first pump down, it may take several days to reach 7e-7 mbar vacuum.

If the issue (not reaching 7e-7 mbar) persists, contact GSS.

Not able to find a solution? Click here to request help.