Z and Zp axis errors on an ACQUITY Common Platform Sample Manager (Guided Troubleshooting) - WKB98387

Article number: 98387

SYMPTOMS

  • Error "Zp Axis Home H/W Fault"
  • Error "Z Axis Home H/W Fault"
  • Error "Z Axis Move H/W Fault"
  • Error "Zp Axis Move H/W Fault"
  • Error "Z axis calibration out of range"
  • Error "Zp axis calibration out of range"
  • Error "Zp axis h/w lost steps"
  • Error "Z axis h/w lost steps"

ENVIRONMENT

  • ACQUITY UPLC H-Class SM-FTN
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY Arc Sample Manager FTN-R

STRATEGY

  1. Ensure that the plate configuration is correctly set.
    1. For Empower - WKB56331
    2. For UNIFI - WKB22166
  2. Look for obstructions to the needle carriage - WKB2119.
  3. Ensure that the needle is not damaged or bent - WKB26140.
  4. Ensure needle does not touch bottom of vial - WKB11021

ADDITIONAL INFORMATION