Error : Z axis home h/w fault - Sample Manager FTN - WKB2119

Article number: 2119

SYMPTOMS

  • Error in the console shows z axis home h/w fault
  • Reboot and reset through console does not resolve
  • Grinding noise when Sample Manager FTN tries to home after reset

 

ENVIRONMENT

  • ACQUITY H Class
  • SM-FTN

CAUSE

Needle carriage obstructed

FIX or WORKAROUND

Perform the following steps :

  1. Remove the two plates.
  2. Power-off the SM-FTN.
  3. Move the needle carriage gently back and forth, being careful not to damage the flex cable.

Z axis home fault.PNG

  1. Power-on SM-FTN.
  2. Replace the plates.
  3. Reset the SM-FTN.