No signal from the 2487 detector seen in Empower - WKB21189

Article number: 21189

SYMPTOMS

  • No signal observed in Empower  
  • Signal is observed from the WFCIII

ENVIRONMENT

  • 2487 Dual Absorbance Detector
  • 2707 Autosampler
  • 1525 Binary HPLC Pump
  • Waters Fraction Collector (WFC III)
  • Empower 3

CAUSE

No injection start cable connected

FIX or WORKAROUND

Connect I/O cable part 441000874 from the 2707 to the 2487.


2707 Capture.JPG

ADDITIONAL INFORMATION

 

id21189, 2487, 2707, 2707SFC