No signal from the 2487 detector seen in Empower - WKB21189

Article number: 21189

SYMPTOMS

  • No signal observed in Empower  
  • Signal is observed from the WFCIII

ENVIRONMENT

  • 2487 Dual Absorbance Detector
  • 2707 Autosampler
  • 1525 Binary HPLC Pump
  • Waters Fraction Collector (WFC III)
  • Empower 3

CAUSE

No injection start cable connected

FIX or WORKAROUND

Connect I/O cable part 441000874 from the 2707 to the 2487.


2707 Capture.JPG

ADDITIONAL INFORMATION

 

id21189, 1525, 1525U, 1525UIVD, 2487, 2707, 2707SFC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP