Z-axis h/w fault, needle breaks during first injection with UNIFI - WKB22166
Article number: 22166
SYMPTOMS
- "Z-axis H/W Lost steps"
- Needle breaks during first injection on UNIFI system
- z-axis h/w fault during first injection on UNIFI system
- Depending on the plate type selected, the needle damage or errors may not show up until the end of the row or column
ENVIRONMENT
- UNIFI
- ACQUITY UPLC H-Class
- ACQUITY UPLC I-Class
- ACQUITY UPLC M-Class
CAUSE
The wrong sample plate configuration is selected in UNIFI by default when a new user is created or UNIFI used for the first time, causing the needle to touch the bottom of the plate or even the wall of the vial.
FIX or WORKAROUND
Select the correct sample plate configuration for the analysis in UNIFI. This must be changed for any new user created in UNIFI upon the first time the user is injecting samples.
In this case, the ANSI - 48 Vial 0.65mL holder was selected instead of the ANSI - 48 Vial 2mL holder.