High backgroud mass 102 (triethylamine) on Xevo G2-XS - WKB221754

Article number: 221754

SYMPTOMS

  • High background mass 102 (triethylamine)

ENVIRONMENT

  • BSM
  • Xevo G2-XS

CAUSE

Triethylamine was added to methonal as the mobile phase, contaminating the tube from degasser port B1 to SSV B.

FIX or WORKAROUND

  1. Avoid mass 102 using another tube (B2)
  2. Replace the tube (430001114).

ADDITIONAL INFORMATION