Xevo TQD tune page is unresponsive in MassLynx - WKB77706
Article number: 77706
SYMPTOMS
- Tune page opens, but buttons do not respond
- Tune page voltage readbacks are zero, and many indicator lights are showing red in diagnostics
- Status in the bottom-right corner says "Instrument Present"
ENVIRONMENT
- MassLynx
- SQ Detector
- SQ Detector 2
- TQ Detector
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-GC
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
- Xevo TQ-S cronos
- Xevo TQ Absolute
CAUSE
Instrument has lost communication with the host PC.
FIX or WORKAROUND
Instrument requires an EPC reboot. There are a few ways to reboot the EPC.
A) Reboot thru Telnet
- Close Masslynx and the MS Console
- Go the START > Windows Systems > Run (note: the Run option may be pinned to the START menu)
- Enter: telnet epc in the Run dialog and hit OK
- At the blinking cursor type reboot and hit enter
- Wait about 2 to 3 minutes for the epc to reboot. You may hear a chime at the end of the boot process
- Open Masslynx and verify the Not Scanning status at the bottom of the sample list page.
B) Reboot using the reset button.
- Close Masslynx and the MS Console
- Go the START > Windows Systems > Run (note: the Run option may be pinned to the START menu)
- Push the reset button on the front of the instrument, often found inside a small hole.
- Wait about 2 to 3 minutes for the epc to reboot. You may hear a chime at the end of the boot process
- Open Masslynx and verify the Not Scanning status at the bottom of the sample list page.
C) Reset both EPC and the IP address for the mass spec
- Close MassLynx and the MS Console.
- Reboot the host PC.
- Remove MS from the Waters DHCP server configuration.
- Reboot the embedded PC.
- Confirm that IP address, MAC address, Name, and Type all repopulate in the DHCP server configuration window.
- Once MassLynx opens again, the status in the bottom-right corner should say "Not Scanning".
ADDITIONAL INFORMATION
The following KCS articles are useful for rebooting the EPC:
https://support.waters.com/KB_Inst/Mass_Spectrometry/WKB7776_How_do_I_reboot_the_EPC_in_my_Mass_Spec
https://support.waters.com/KB_Inst/Mass_Spectrometry/WKB7610_No_communications_with_the_XEVO_TQD
