Skip to main content
Waters

How do you power-off a Xevo TQD in preparation for a scheduled power outage? - WKB7770

Article number: 7770

ENVIRONMENT

  • Xevo TQD

ANSWER

  1. Open the MS Tune page.
  2. Place the instrument into Standby mode and turn off all gases.
  3. Go to the Vacuum menu and select VENT. Wait for the pumps to turn off.
  4. Close MassLynx and the ACQUITY MS Console.
  5. Power-off the Xevo TQD.
  6. Power-down the PC.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.