"Server Busy" Error stopping MassLynx during overnight runs - WKB3249
Article number: 3249
SYMPTOMS
- MassLynx sample list acquisition pauses randomly
- Run pauses during overnight run
- Operating system "Server Busy" error shows on pop-up window
- No other MassLynx error messages
ENVIRONMENT
- MassLynx
- All MS systems
- Third-party software
- Antivirus software (e.g. Bitdefender, Sophos, Eset Node32, McAfee, Symantec, Kaspersky, Cortex XDR)
CAUSE
Server Busy messages are caused by resource conflicts. Usually third-party programs using resources that MassLynx is trying to use.
This can be antivirus software (Sophos, Eset Node32, McAfee, Symantec), file transfer programs/backup software (Avamar), or applications performing automatic updates (Google Toolbar).
FIX or WORKAROUND
Remove the resource conflict by doing the following:
- Set the antivirus software to exclude all MassLynx related processes.
- Only allow full antivirus scans when MassLynx is not acquiring data.
- Only allow backup software to run when MassLynx is not acquiring data.
- Disable programs' automatic updates during MassLynx acquisition.
ADDITIONAL INFORMATION
See also articles:
Recommended antivirus exclusions, firewall setup and UAC settings for MassLynx - WKB21150
How to configure Symantec Antivirus software for MassLynx
Server Busy Error MassLynx