In MassLynx, the shutdown method fails to run to completion - WKB32249

Article number: 32249

SYMPTOMS

  • System should shut down on error - e.g. LC shows missing vial or wrong plate error - but shutdown method does not stop the LC flow
  • MS may be found in Standby mode
  • The shutdown method entry in the shutdown logs states that it failed to load the LC method 

ENVIRONMENT

  • MassLynx 4.1
  • Xevo G2-S

CAUSE

The shutdown method is using an LC method from a different project.

FIX or WORKAROUND

Ensure that the LC method needed is in the current MassLynx project.

ADDITIONAL INFORMATION

 