In MassLynx, the shutdown method fails to run to completion - WKB32249
Article number: 32249
SYMPTOMS
- System should shut down on error - e.g. LC shows missing vial or wrong plate error - but shutdown method does not stop the LC flow
- MS may be found in Standby mode
- The shutdown method entry in the shutdown logs states that it failed to load the LC method
ENVIRONMENT
- MassLynx 4.1
- Xevo G2-S
CAUSE
The shutdown method is using an LC method from a different project.
FIX or WORKAROUND
Ensure that the LC method needed is in the current MassLynx project.
ADDITIONAL INFORMATION