Results not displayed in the FractionLynx Browser with SCN 1050 - WKB322650

Article number: 322650

SYMPTOMS

  • Nothing is displayed in the FractionLynx Browser when a results file is opened in SCN 1050
  • No data can be seen displayed but it is there when the report is printed
    FractionBrowserBeforePlateConfiguration.JPG
  • The printed report shows results but the browser is empty with no results showing

ENVIRONMENT

  • FractionLynx
  • MassLynx

CAUSE

A bug in SCN 1050 means that results are not displayed unless the plate is configured in the browser - the default is X,Y Numeric, Numeric. 
DefaultFractionLynxPlateDisplay.JPG
Changing this to anything other than X,Y Numeric, Numeric - e.g. Sequential Discontinuous or X,Y Letter, Number will change the display so the results can be shown

CRI-9337 was raised. 

FIX or WORKAROUND

  1. Change the plate display option in the FractionLynx Browser
  2. In FractionLynx Browser click View then Options
  3. Select the Default Plate tab
  4. Under Grid Reference change the display to match the plate in use
  5. Or change the plate from Numeric Numeric to any other display - e.g. Sequential Discontinuous, or X, Y Numeric, Letter
    FractionLynxPlateSequentialDiscontinuous.JPG
  6.   This will be fixed in SCN 1054.

ADDITIONAL INFORMATION