QDa Windows 10/11 setup in Empower (Guided Troubleshooting) - WKB99782
Article number: 99782
SYMPTOMS
- Waters Computer Images for Windows 10 have local group policy that subjects users with administrator privileges to User Account Control; this results in a variety of symptoms, including improper instrument control and problems installing, running, and configuring software
- Console cannot find QDa
- Console finds QDa but displays communication failure; unable to reset
- QDa applet in Run Samples shows MS Instrument and Communication Error
- DHCP Server shows QDa
- Node Properties Shows QDa as "Yes"
- Instrument configures normally in Empower
- In a networked Empower Enterprise system, the ACQUITY QDa does not work on a LAC/E with a bridged network card
- System stuck at Preparing for Injection during acquisition and never completes the acquisition
ENVIRONMENT
- ACQUITY QDa Mass Detector
- Windows 10
- Empower Enterprise
STRATEGY
- How to disable fast startup on Windows 10 - WKB53796
- How to set Windows firewall settings for Empower - WKB47333
- How to bridge network cards in Windows 10 - WKB78169
- Which version of Empower supports Windows 10? - WKB27485
- How to install and enable Telnet in Windows 7 and 10 - WKB5625
- How to disable IPv6 - WKB9503
- How to apply the Windows firewall settings to just one network card - WKB58216
- QDa does not work on LAC/E bridged network card in an Empower Enterprise system (INFHW-197, CRI-1854) - WKB95041