Troubleshooting LockSpray signal issues on QTofs and Tofs with IntelliStart (Guided Troubleshooting) - WKB96098

Article number: 96098

SYMPTOMS

  • Signal issues via reference probe
  • No signal
  • Low signal intensity
  • Unstable signal
  • Delayed flow/signal
  • Decaying signal

ENVIRONMENT

  • Xevo QTof
  • Xevo G2 
  • Xevo G2-S
  • Xevo G2-XS
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • IntelliStart

STRATEGY

  1. Use higher flow rate - WKB13748
  2. Ensure that orange fluidics tubing is fully submerged in the wash solution - WKB69940
  3. Most Likely:  Purge reference fluidics (at least 5 times) - WKB92145
  4. Tighten PEEK nut at end of the reference probe mount assembly - WKB57869
  5. Trim the end of or replace the tubing from reservoir B to the reference valve - WKB51661
  6. Use a more aqueous LockSpray solution - WKB13748
  7. Rebuild reference probe - WKB6834
  8. Purge syringe pump and blue tubing directly with methanol - WKB67224