Troubleshooting LockSpray signal issues on QTofs and Tofs with IntelliStart (Guided Troubleshooting) - WKB96098
Article number: 96098
SYMPTOMS
- Signal issues via reference probe
- No signal
- Low signal intensity
- Unstable signal
- Delayed flow/signal
- Decaying signal
ENVIRONMENT
- Xevo QTof
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- IntelliStart
STRATEGY
- Use higher flow rate - WKB13748
- Ensure that orange fluidics tubing is fully submerged in the wash solution - WKB69940
- Most Likely: Purge reference fluidics (at least 5 times) - WKB92145
- Tighten PEEK nut at end of the reference probe mount assembly - WKB57869
- Trim the end of or replace the tubing from reservoir B to the reference valve - WKB51661
- Use a more aqueous LockSpray solution - WKB13748
- Rebuild reference probe - WKB6834
- Purge syringe pump and blue tubing directly with methanol - WKB67224
ADDITIONAL INFORMATION