No signal on LockSpray beam for QTof using fluidics line B - WKB51661
Article number: 51661
SYMPTOMS
- No LockSpray signal when using line B on the reference probe
- Good signal on sample spray
- Change of LockSpray solution to line A or C brings good signal again
ENVIRONMENT
- Xevo QTof
- Xevo Tof
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S
- Xevo G2-XS
CAUSE
Blockage on the reference tubing on line B going from the solution to the reference valve
FIX or WORKAROUND
- Replace the reference tubing for line B from solution to the reference valve. See Parts locator.
- If the blockage occurred on the end of the tubing, use the PEEK cutter to cut the tip of the tubing.
- Reconnect the tubing on the reference valve.
ADDITIONAL INFORMATION
id51661, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSTF, XEVOQTOF