Low lockspray sensitivity on Xevo QTof when doing LC-MS runs - WKB13748
Article number: 13748
SYMPTOMS
- LockSpray probe does not have a blockage or a leak, yet spray is unstable
ENVIRONMENT
- Xevo G2-S Q-Tof
- Xevo G2-XS QTof
CAUSE
The high heat from the sample sprayer desolvation heater and the high-organic solvent cause the lockspray solvent to boil off, resulting in an unstable beam.
FIX or WORKAROUND
Go to a higher percentage of aqueous in the lockspray solution (approximately 80% H2O) and/or go to a higher flow rate.