Trap current limiting problem on the AutoSpec 1 - WKB96173

Article number: 96173

SYMPTOMS

  • The EI trap regulating current may regulate correctly but then drops shortly after; the filament current increases to maximum, and the emission current increases; the electron energy remains constant at the set value
  • CI control of the source/filament works correctly

ENVIRONMENT

  • AutoSpec Premier
  • AutoSpec Ultima

CAUSE

Poor connection on the outer source trap wiring.

FIX or WORKAROUND

  1. The wire connecting the Outer Source trap contact to the Contact Rail, left-hand rail - contact 4, is loose
  2. Verify push fit of the Cambion connector on the contact ring, LHS positon 4, underneath the outer source assembly.
  3. Check the M1.6 screw/nut connection securing the copper wire to the trap contact itself. It should be tight with no play in the connection.  
    • Tighten as necessary with care. Small thread, 1.6 mm. Nut size across the flats, 3.2 mm.   

ADDITIONAL INFORMATION

See related articles WKB8643, WKB11064, WKB20704, WKB96181, WKB124360 and WKB124362

See also WKB62609 for conditioning filaments and WKB62805 for tips on increasing their lifetime.