Unstable LockSpray beam on QTof - WKB6834
Article number: 6834
SYMPTOMS
- LockSpray beam is sporadic
- Low reference signal
- Increasing the flow rate improves the stability of the LockSpray beam
- Dropping the desolvation temperature does not help stabilize the LockSpray beam
- No clicking heard from the LockSpray fluidics
ENVIRONMENT
- Xevo QTof
- Xevo G2-S
- Xevo G2-XS
CAUSE
There is a blockage on the LockSpray probe.
FIX or WORKAROUND
- Rebuild the LockSpray probe.
ADDITIONAL INFORMATION
Log into the Waters website and click here to search for the Overview and Maintenance Guide, which describes how to rebuild the LockSpray probe for your particular QTof.
