Xevo TQD areas are inconsistent - WKB81574
Article number: 81574
SYMPTOMS
- Decreased sensitivity
- Retention times are consistent
ENVIRONMENT
- Quattro Premier
- Quattro Premier XE
- Quattro micro API
- Quattro Ultima
- 3100 Mass Detector
- ACQUITY SQD system
- SQ Detector
- ACQUITY TQD system
- TQ Detector
- SQ Detector 2
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
- ZQ 2000
- ZQ 4000
CAUSE
Dirty sample cone
FIX or WORKAROUND
Clean the sample cone.
ADDITIONAL INFORMATION
See article WKB10984 for directions on cleaning the sample cone.
id81574, A-SQ, A-SQ, MLYNX, MLYNXV41, QUATPREMXE, QUMICRO, QUPREMIER, SQD, SQD2, SUPMM, TQD, TQDIVD, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD, ZQ2000, ZQ2000SFC, ZQ4000