No beam because of communication issue with EPC - WKB81251
Article number: 81251
SYMPTOMS
- No beam, not even solvent beam.
- All voltage readbacks are normal.
- Desolvation temperature option was unavailable and showed readback of 40 °C.
ENVIRONMENT
- Quattro Premier XE
- Xevo TQs Micro
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
- MassLynx
- UNIFI
CAUSE
Communication issue with EPC.
FIX or WORKAROUND
Restart EPC (see detailed instructions here - WKB7776).
ADDITIONAL INFORMATION
For instruments under UNIFI control, reset the LND to regain communication. If that fail to restore communication restart the LND and perform a hard reboot (electronics reboot) of the MS.
After EPC restart, desolvation temperature box returned to normal and all beams were visible.
How to reboot the EPC on a ZQ or Quattro micro - WKB66315