How to physically reboot the EPC on a ZQ or Quattro Micro - WKB66315

Article number: 66315

OBJECTIVE or GOAL

Rebooting the EPC on a ZQ or Quattro Micro physically

ENVIRONMENT

  • ZQ 2000
  • ZQ 4000
  • Quattro micro API
  • Quattro Micro GC

PROCEDURE

  1. Locate the unlabeled hole in the front panel, approximately 3/4 of the distance down from the diverter valve to the syringe drive.
  2. Use PEEK tubing or something similar to push the EPC reset button inside this hole.
  3. Confirm that the EPC has rebooted successfully by connecting to MassLynx.

ADDITIONAL INFORMATION

id66315, QUMICRO, QUMICROGC, ZQ2000, ZQ2000SFC, ZQ4000