Blown F19 fuse continues to allow the syringe pump to function - WKB66279

Article number: 66279

SYMPTOMS

  • Blown F19 fuse continues to allow the syringe pump to function
  • F19 fuse remains in Red state on Tune Page Diagnostic
  • Replaced F19 and powered-on instrument; F19 shows blown at startup and is blown via DVM

ENVIRONMENT

  • Xevo TQ-S micro
  • Xevo TQ-S

CAUSE

Faulty personality PCB

FIX or WORKAROUND

Install a new personality PCB.

ADDITIONAL INFORMATION

 