Skip to main content
Waters

How to start up or pump down a Waters triple quad - WKB5462

Article number: 5462

OBJECTIVE or GOAL

Pump down a triple quad from tune page

ENVIRONMENT

  • MassLynx 4.1
  • Empower
  • MassLynx 4.2
  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQD
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • Xevo TQ-GC
  • TQ Detector
  • SQ Detector
  • SQ Detector 2
  • Xevo TQ Absolute

PROCEDURE

  1. Power-on the host PC, and then log in to Windows.
  2. Power-on the LC modules and let it initialize.
  3. Power-on the mass spectrometer: For a Xevo TQ-S, TQ MS, TQ-XS or Absolute, ensure that the MS power cord is plugged in and that both toggle switches on the back (middle-bottom) are in the up position. For all other triple- or single-quads, power on the MS using the power button .
  4. Wait five minutes for the embedded PC (EPC) to boot up.
  5. For Masslynx- Go into MassLynx and open the tune page, For Empower- Go to empower sample set and open the instrument console, Go to Mass spec tab and open the tune page. 
  6. From the tune page, select the vacuum (top drop-down menu in the middle on the tune page), and select pump. The MS will pump down faster if the isolation valve is closed. Remember to open valve before going into Operate mode (this is critical if you have an S or XS).
  7. To pump down faster, on tandem quads turn the collision gas on. The source (ion block) isolation valve can be closed to assist in pump down, however, it must be opened before going into Operate.
  8. Once the MS pumps for an hour, you can put it into Operate mode. However, for optimum sensitivity (as a result of good vacuum), it is best to wait until the next day to run samples.

ADDITIONAL INFORMATION

If you have an H-Class or I-Class, power on the Sample Manager first.

 

See How to shut down a Xevo TQ-S.

id5462, MLYNX, MLYNXV41, SUPMM, TQD, TQDIVD, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD

Not able to find a solution? Click here to request help.