Loss of comunication due to Firewall ON in local area conections . - WKB25516
Article number: 25516
SYMPTOMS
- Loss of communication.
- In WDHCP configuration server, UPLC modules and XEVO TQ-S with IP, but there are no recognized modules or serial numbers.
ENVIRONMENT
- Xevo TQ-S
- Xevo TQ-XS
- Lenovo
- ACQUITY UPLC
- BioAccord System
- ACQUITY RDa Detector
CAUSE
Firewall ON in all local area connections (local area connection 1 and local area connection 2).
Local area connections in the same PC´s mother board.
FIX or WORKAROUND
An extra network card (LAN). Windows Firewall OFF for this extra card and firewall ON for the card in the mother board.