The Xevo TQ-XS Qualification Workbook has incorrect m/z values referenced in the maintenance procedure - WKB255162
Article number: 255162
SYMPTOMS
- Incorrect m/z values referenced in Waters qualification document 715005179 Rev B
- Pages 10 and 14 of qualification document 715005179 Rev B reference m/z values of 59.1, 455.3,1080.8, and 2034.6, which are incorrect for the Xevo TQ-XS setup solution
ENVIRONMENT
- Xevo TQ-XS
- Qualification workbook
CAUSE
Qualification workbook maintenance procedures are no longer updated per service alert 715008484.
FIX or WORKAROUND
- Service Alert Document 715008484 indicates to use the applicable PM protocols in place of the Qualification Workbook Maintenance Procedures (MP).
- The latest PM Protocol revision of 715005520 should be used, and the correct m/z values 95, 455.3, 1121.9, and 2017.6 should be used.
ADDITIONAL INFORMATION
id255162, UNIFQLIC, XEVOG2XSQT, XEVOTQXS, XEVTQXSIVD