Active preheater leaks - WKB94209

Article number: 94209

SYMPTOMS

  • Leak found between active preheater and column
  • Abnormal chromatography peak shape
  • Shifted retention time of peaks

ENVIRONMENT

  • ACQUITY UPLC Column Manager
  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS

CAUSE

The active preheater fitting has been overtightened to the column. This can distort the PEEK ferrule, causing a dead volume at the head of the column and leading to leaks or abnormal chromatography.

FIX or WORKAROUND

  1. Replace the PEEK ferrule on the active preheater.
  2. Avoid overtightening the active preheater fitting on the column.

ADDITIONAL INFORMATION

How-to video about Install 3-Piece Finger-Tight Fitting with APH

The part number is 700003114 - 10/pk Ferrule (PEEK).

700010009 - Kit, Ferrule, Back Locking, PEEK, 1/16 (10/pk) or 700011767 - Kit, Ferrule Back Locking PEEK 1/16 (2/pk)