Cannot put PEEK column ferrule onto the Active Preheater tubing - WKB7194

Article number: 7194

SYMPTOMS

  • Replacing column
  • Installing new ferrules onto the Active Preheater
  • Ferrule will not slide onto tubing
  • Tubing appears damaged, no longer round

ENVIRONMENT

  • ACQUITY UPLC I-Class Sample Manager FTN - FL
  • ACQUITY UPLC I-Class Column Heater - Active (CH-A)
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY APC

CAUSE

The end of the tubing on the Active Preheater is damaged.  An old PEEK ferrule was tightened too many times with a wrench. Repetitive over-tightening of the ferrule has warped the metal tubing of the active preheater, causing it to be too wide for the new ferrule to fit.

clipboard_e295998fd3965f9a2c78be0330b039f91.png

FIX or WORKAROUND

Replace the Active Preheater.

ADDITIONAL INFORMATION

The PEEK ferrule is designed for only a few uses per ferrule.  Ferrules should be changed out if leaking is observed when installing a column.

ACQUITY I-Class Active Pre-Heater (APH, SS, 12.5 LG, 0.003 ID) Part Number 205000894

ACQUITY H-Class Active Pre-Heater (APH, SS, 12.5 LG) Part Number 205000730

ACQUITY Active Pre-Heater PEEK ferrule 1/16" (10/pk) Part Number 700003114

ACQUITY I-Class Column Heater - Active, Active Pre-Heater part number = 205000730  (ACQUITY UPLC APH SS 12.5LG)