Leak under primary check valve on 2695 - WKB71953
Article number: 71953
SYMPTOMS
- Leak under primary check valve
ENVIRONMENT
- 2695 Separations Module
CAUSE
Leak between the GPV inlet tubing and the PPT check valve housing
FIX or WORKAROUND
- Replace the GPV inlet tubing (Standard Flow: WAT270986 / Low Flow: WAT270976 / High Flow: WAT270968).
- Replace the check valve housing (700002332).
ADDITIONAL INFORMATION