How do I confirm that the 2998 Photodiode Array Detector is still within specification? - WKB7196
Article number: 7196
OBJECTIVE or GOAL
Verify wavelength calibration.
ENVIRONMENT
- 2998 Photodiode Array Detector, PDA
- Empower all versions
- 2998 ICS v1.30
PROCEDURE
-
Prepare flow cell and flush with an appropriate intermediate solvent before switching to 100% methanol.
-
From the console, select 2998 PDA Detector from the system tree.
-
Select Maintain > Verify Calibration > Start.
-
When complete, the results pane appears.
ADDITIONAL INFORMATION
The 2998 PDA detector automatically runs a series of internal diagnostic tests upon start-up. Wavelength verification is one of the tests run, so if both LEDs are green, the diagnostics passed.