ACQUITY TUV detector exhibits unacceptable baseline drift - WKB94252
Article number: 94252
SYMPTOMS
- The baseline drift on TUV detector was more than 0.012 AU in 20 minutes, exceeding the qualification standard of 0.01 AU/hour.
- Nothing abnormal was detected when checking the optical path. The lamp energy was acceptable. And the problem remains after flow cell replaced.
- The baseline drift was improved after system inspection performed and detector inlet tubing replaced, but was still unacceptable.
- By monitoring the baseline overnight, it was found that abrupt baseline drift showed up with air conditioner turned off; when the air conditioner was turned on to control the temperature, the baseline gradually became stable and met the instrument parameter and user requirements.
ENVIRONMENT
- ACQUITY UPLC Tunable UV Detector
- ACQUITY UPLC H-Class
- ACQUITY UPLC I-Class
CAUSE
The abrupt baseline drift was due to unstable ambient temperature in the laboratory when the air conditioner was turned off.
FIX or WORKAROUND
- Stabilize the ambient temperature in the laboratory by turning on the air conditioner, the baseline should gradually become flat.
ADDITIONAL INFORMATION
