Xevo TQD MS1 sensitivity is too low to see MS2 - WKB94336
Article number: 94336
SYMPTOMS
- Xevo TQD MS1 sensitivity is too low to see MS2
- Peak sensitivity suddenly decreases in MRM measurement; a cold reset of MS temporarily improves the situation
ENVIRONMENT
- Xevo TQD
- MassLynx
CAUSE
Defective RF2 generator or defective personality board
FIX or WORKAROUND
- Check the rectified voltage and make a voltage division.
- If the rectified voltage of MS2 is 0 V even when confirming with Align mode: Change the MS number to 0-2000 with Span 0 in Normal mode. If the rectified voltage of MS2 is still 0 V, there is no output from the RF2 generator and it needs to be replaced.
- If the MS2 rectified voltage is normal (-3.37 V and so on) when confirming with Align mode: Change the MS number to 0-2000 with Span 0 in Normal mode. If the rectified voltage of MS2 is still 0 V, the MS program on the personality board is not working and the board needs to be replaced.
ADDITIONAL INFORMATION
Normally, MS2 rectified voltage is -3.37 in Align mode, and it varies when MS number is 0-2000 with Span 0 in Normal mode (-4.91 for 1000 mass, -9.85 for 2000 mass).