2545 high delta pressure - WKB94160
Article number: 94160
SYMPTOMS
- High delta pressure
- 2545 static leak test shows low leak rate, but indicates that outlet valve for one of the pump heads fails to close
- When one pump head compresses, pressure increases equally on the other pump head (should stay close to 0psi with properly functioning outlet valves)
ENVIRONMENT
- 2525/2545 Binary Gradient Module
- 2535/2545/2555 Quaternary Gradient Module
CAUSE
Defective outlet check valve.
FIX or WORKAROUND
- Replace the outlet check valve, part number 700001493
ADDITIONAL INFORMATION
id94160, 2545, 2545Q, checkvalve