2545 high delta pressure - WKB94160

Article number: 94160

SYMPTOMS

  • High delta pressure
  • 2545 static leak test shows low leak rate, but indicates that outlet valve for one of the pump heads fails to close
  • When one pump head compresses, pressure increases equally on the other pump head (should stay close to 0psi with properly functioning outlet valves)

ENVIRONMENT

  • 2525/2545 Binary Gradient Module
  • 2535/2545/2555 Quaternary Gradient Module

CAUSE

Defective outlet check valve.

FIX or WORKAROUND

  1. Replace the outlet check valve, part number 700001493

ADDITIONAL INFORMATION