Skip to main content
Waters

ACQUITY SM-FTN system over pressure - WKB6394

Article number: 6394

SYMPTOMS

  • ACQUITY System with SM-FTN reports HW over pressure on solvent manager during run
  • Needle seal readiness test passes with high pressure difference (for example, 1000 psi at 0.1 mL/min) 
  • High pressure at low flow rate (for example, 0.1 mL/min) with the column disconnected going directly to waste

ENVIRONMENT

  • ACQUITY UPLC I-Class SM-FTN
  • ACQUITY UPLC H-Class SM-FTN
  • ACQUITY Premier System

CAUSE

Debris or particulates were forced into the needle seat port or needle seal, causing a blockage and system overpressure. Inspect sample vials and needle wash for particulates or growth. A blockage in this area is nearly always introduced from the injector area. 

FIX or WORKAROUND

  ** Try a reverse flow and flush the seat port first to unclog if it doesn't work then proceed with replacing these parts ** - see WKB115842

  • Replace the needle seal (Vespel with anti-rot seat, part 405011492). 
  • Replace the seat port if required.
  • Ensure that pressure is back to normal and that values are as expected.
  • Run "Characterize Needle Seal".
  • Run "Needle Seal Readiness Test" as a leak test.

ADDITIONAL INFORMATION

Seat port assembly parts:

Before replacing any part, determine whether the injection valve or needle is clogged. See article WKB19544.

id6394, A-30SM, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO

Not able to find a solution? Click here to request help.