SM-FTN error: Column active preheater is not supported with current configuration - WKB63907
Article number: 63907
SYMPTOMS
- Error in ACQUITY UPLC Sample Manager
Column active preheater is not supported with current configuration - Resetting does not resolve
ENVIRONMENT
- ACQUITY UPLC Sample Manager-Flow Through Needle
- ACQUITY UPLC Column Manager
- 30-cm Column Heater
CAUSE
The active pre-heater is not connected to the 30 cm column heater, while the active pre-heater configuration in FTN Instrument Method is "enable"
FIX or WORKAROUND
Set the active pre-heater to "disable” in FTN Instrument Method.
ADDITIONAL INFORMATION
id63907, A-30SM, APH, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO