Skip to main content
Waters

SM-FTN: How to back flush a blockage in Vespel seal, seat port tubing or sample needle - WKB115842

Article number: 115842

SYMPTOMS

  • Overpressure or higher pressure than normal when injection valve is in "Inject" position
    clipboard_e6ab4bd1a409ab238217e3fa2685dcc23.png
  • "Sample fluidics high pressure limit" error seen in the ACQUITY Console
  • "Draw rate excessive" error seen in the ACQUITY Console
  • High pressure error is seen even when a column is not installed

ENVIRONMENT

  • ACQUITY UPLC H-Class SM-FTN
  • ACQUITY UPLC I-Class SM-FTN
  • ACQUITY Arc Sample Manager FTN-R

CAUSE

 Blockage in sample needle, needle seal or seat port tubing.

FIX or WORKAROUND

clipboard_d1e0cf4c-b592-407a-b253-610671bb28e6.png

  1. Open the sample manager door and reset using ACQUITY console. The SM-FTN will move the injection valve to "Load" position and moves the needle carriage to the rear of the sample compartment. You may need to close the door to stop the cyclic resetting and complete the needle movement away from the wash station. The important point is that the needle is not in the injection port / wash station during the back flushing. The sample manager door must remain open during this diagnostic flushing procedure.
  2. Using a 1/4" spanner, disconnect port 6 (outlet to column) and disconnect port 4 (sample needle) on the SM-FTN injection valve.
  3. Connect the sample needle (from port 4) to port 6 and using a suitable mobile phase for your sample, set a low flow rate on the solvent manager (0.1 mL/min) via ACQUITY console to flush needle and monitor for a significant drop in pressure that could indicate that a blockage has been cleared.  A higher flow rate and pressure may be needed to clear the blockage.  If no high pressure is seen during this flushing step, then continue to step #4.
    NOTE: mobile phase will drip into the sample compartment from the needle so place a small amount of absorbent tissue inside to absorb the liquid.
  4. Repeat for seat port tubing and vespel seat by connecting port 1 (needle seat tubing) to port 6 (outlet to column) and flush as previously described above.  No drips will be seen during this flushing step as the liquid is contained within the needle wash station.  Monitor system pressure for a significant drop in pressure that could indicate that a blockage has been cleared. A higher flow rate and pressure may be needed to clear the blockage.  If no high pressure is seen during this flushing step, then there might be an issue with injector valve pod and it should be replaced.
  5. Return tubing to original configuration, remove any absorbent tissue from inside the sample compartment and check system pressures by perform a injection with a blank / solvent sample.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.