Skip to main content
Waters

SM-FTN: How to back flush a blockage in Vespel seal, seat port tubing or sample needle - WKB115842

Article number: 115842

SYMPTOMS

  • Overpressure or higher pressure than normal when injection valve is in injection position
  • Sample fluidics high pressure limit
  • Draw rate excessive errors in the ACQUITY Console

ENVIRONMENT

  • ACQUITY UPLC H-Class SM-FTN
  • ACQUITY UPLC I-Class SM-FTN
  • ACQUITY Arc SM-FTN-R

CAUSE

 Blockage in sample needle, needle seal or seat port tubing.

FIX or WORKAROUND

  1. Open the sample manager door and reset using Acquity console. The SM-FTN will move the injection valve to the "Load" position and locates the needle carriage to the rear of the sample compartment. You may need to close the door to stop the cyclic resetting and complete the needle movement away from the wash station. The important point is that the needle is not in the injection port / wash station during the back flushing.
  2. Using a 1/4" spanner, disconnect port 6 (outlet to column) and disconnect port 4 (the sample needle) on the SM-FTN injection valve.
  3. Connect the sample needle to port 6 and using a suitable mobile phase for your sample, set a low flow rate on the solvent manager (0.1 mL / min) via Acquity console to flush needle and look for a significant drop in pressure that could indicate that a blockage has cleared. NOTE: mobile phase will drip into the sample compartment from the needle so place a small amount of absorbent tissue inside.
  4. Repeat for seat port tubing / vespel seat by connecting port 1 (needle seat tubing) to port 6 (outlet to column) and flush as previously described above.
  5. Return tubing to original configuration, remove any absorbent tissue from inside the sample compartment and check system pressures by perform a injection with a blank / solvent sample.

ADDITIONAL INFORMATION

 

id115842, eluent, UPARBSMFTN, UPARHSMFTN, UPARPFTN, UPARSMFTN

Not able to find a solution? Click here to request help.