Needle/seal move limit exceeded - WKB115820

Article number: 115820

SYMPTOMS

  • Needle/seal move limit exceeded

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS
  • SM-FTN
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio

CAUSE

Motor driver PCB

FIX or WORKAROUND

Replace the motor driver PCB (700005228).

ADDITIONAL INFORMATION