How to adjust the needle depth on a 2695 Separations Module - WKB63989

Article number: 63989

OBJECTIVE or GOAL

Adjust the needle depth for use of nonstandard vials, particularly low-volume inserts.

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

PROCEDURE

From the front panel of the 2695, select Diag > Service Utilities > Custom Offset and specify the value listed in Table B.4 of the 2695 Separations Module Operator's Guide.

ADDITIONAL INFORMATION

id63989, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB