ACQUITY UPLC - sample syringe leak test fails - WKB49303

Article number: 49303

SYMPTOMS

  • Sample syringe leak test fails after priming
  • No air visible in sample syringe upon priming
  • No damage on 2 or 3 port tubing of injector valve

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - FTN
  • ACQUITY APC Sample Manager - FTN
  • ACQUITY Arc Sample Manager FTN-R
  • ACQUITY I-Class Sample Manager - Fixed Loop
  • ACQUITY I-Class Sample Manager - FTN

CAUSE

The sample syringe plunger seal leaks.

FIX or WORKAROUND

  1. Replace the sample syringe (50 µL = 410001348; 100 µL = 700002570; 250 µL = 410001347; 500 µL (HPLC) = 410001647, APC - 100ul 70009529)
  2. Verify functionality by performing sample syringe leak test successfully
  3. When in use a leaking sample syringe can be inspected to see if solvent present by the plunger.

ADDITIONAL INFORMATION

WKB4005 SM-FTN syringe leak test fails intermittently

WKB54861 Sample manager FTN leak test fails

WKB56424 Failing sample syringe leak test on Sample Manager FTN

WKB57007 Sample syringe leak test fails on ACQUITY sample manager

WKB62299 ACQUITY Sample Manager FTN failed syringe leak test after PM

id49303, A-30SM, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARQSM, UPARSMFTN, UPBSMIVD, UPIBSMIVD, UPISMFTN, UPLCCOLHTR, UPNBINARY, UPPDAARCB, UPPFTN, UPSMFTN, UPSMFTN+, UPSMIVD, UPSMPMGR