SM-FTN syringe leak test fails intermittently - WKB4005

Article number: 4005

SYMPTOMS

  • SM-FTN syringe leak test fails after PM

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class
  • Sample Manager FTN

CAUSE

Air in the fluidics lines

Leak at the sample transducer.

FIX or WORKAROUND

  • Purge thoroughly
  • Tighten all fittings on the transducer.
  • If necessary cut peek tubing with peek cutter

ADDITIONAL INFORMATION

LeakingTransducer.png

 