SM-FTN syringe leak test fails - WKB4005

Article number: 4005

SYMPTOMS

  • SM-FTN syringe leak test fails
  • Carry-over
  • Poor reproducibility

ENVIRONMENT

  • ACQUITY UPLC H-Class Sample Manager FTN
  • ACQUITY UPLC I-Class Sample Manager FTN

CAUSE

Air in the fluidics lines. Bubble(s) seen in the syringe.

FIX or WORKAROUND

  • Prime the SM-FTN with both wash solvent and purge solvent. Prime the wash solvent at least 60 seconds and the purge solvent at least 10 times.
    • While priming the purge solvent verify that there are no bubbles in the syringe.
    • If there are bubbles, flick the metal pin with a finger several times until all bubbles are removed.

ADDITIONAL INFORMATION

Sample Syringe leak test can fail directly after PM.

Transducer with fittings, tighten all in case that a leak has caused the air in the fluidics. If not able to get rid of leak, cut the ends of the PEEK tubing with PEEK cutter and tighten.

LeakingTransducer.png

 

 

id4005, A-30SM, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO