Skip to main content
Waters

SM-FTN syringe leak test fails - WKB4005

Article number: 4005

SYMPTOMS

  • SM-FTN syringe leak test fails
  • Carry-over
  • Poor reproducibility

ENVIRONMENT

  • ACQUITY UPLC H-Class Sample Manager FTN
  • ACQUITY UPLC I-Class Sample Manager FTN

CAUSE

Air in the fluidics lines. Bubble(s) seen in the syringe.

FIX or WORKAROUND

  • Prime the SM-FTN with both wash solvent and purge solvent. Prime the wash solvent at least 60 seconds and the purge solvent at least 10 times.
    • While priming the purge solvent verify that there are no bubbles in the syringe.
    • If there are bubbles, flick the metal pin with a finger several times until all bubbles are removed.

ADDITIONAL INFORMATION

Sample Syringe leak test can fail directly after PM.

Transducer with fittings, tighten all in case that a leak has caused the air in the fluidics. If not able to get rid of leak, cut the ends of the PEEK tubing with PEEK cutter and tighten.

LeakingTransducer.png

Not able to find a solution? Click here to request help.