ACQUITY UPLC H-Class/ACQUITY UPLC I-Class SM-FTN: Drawing Sample Rate Excessive - WKB399
Article number: 399
SYMPTOMS
- Drawing sample rate excessive message is a user warning and does not stop the system from continuing with an injection or purge wash.
- It is considered a caution and recorded within the console log.
ENVIRONMENT
- ACQUITY UPLC H-Class Sample Manager - Flow Through Needle (SM-FTN)
- ACQUITY UPLC I-Class Sample Manager - Flow Through Needle (SM-FTN)
CAUSE
Drawing sample rate excessive can be caused by the syringe draw rate being too fast, possibly because the solution is viscous.
FIX or WORKAROUND
- Define the optimum speed of the syringe to reduce the occurrence of the message appearing
- Decrease the syringe draw rate in the instrument or inlet method.
ADDITIONAL INFORMATION
- The "drawing sample rate excessive" message may possibly be a normal occurrence for the application
- Verify the correct sample needle, extension loop and syringe is configured and installed.
- The pathway is from the sample syringe, pressure transducer, metering syringe valve and the injector/pod cartridge (port position 3 and 4) and through the needle.
- 10uL/min is the lowest you can go, so start with that .
- Make sure the needle is not touching the bottom of the vial. You can raise the needle to 4mm and slow the syringe speed to 10uL/min, then you can lower the needle back and see what fixed it.
id399, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO