Sample syringe leak test fails on ACQUITY UPLC Sample Manager-FTN - WKB56424

Article number: 56424

SYMPTOMS

  • SM-FTN fails sample syringe leak test
  • No peaks in chromatogram

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Flow Through Needle
  • ACQUITY UPLC Sample Manager

CAUSE

Air trapped in syringe fluidics

FIX or WORKAROUND

  1. For SM-FTN, detach the outlet check valve from the injection valve (port 2), reprime the syringe, and then reattach the check valve.
  2. Also, try to tighten the sample syringe counter-clockwise.
  3. Repeat the leak test.