Sample syringe leak test fails on ACQUITY UPLC Sample Manager-FTN - WKB56424
Article number: 56424
SYMPTOMS
- SM-FTN fails sample syringe leak test
- No peaks in chromatogram
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Flow Through Needle
- ACQUITY UPLC Sample Manager
CAUSE
Air trapped in syringe fluidics
FIX or WORKAROUND
- For SM-FTN, detach the outlet check valve from the injection valve (port 2), reprime the syringe, and then reattach the check valve.
- Also, try to tighten the sample syringe counter-clockwise.
- Repeat the leak test.
ADDITIONAL INFORMATION
id56424, A-30SM, checkvalve, UPBSMIVD, UPIBSM, UPIBSM+, UPIBSMIVD, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMIVD, UPSMPMGR, UPSMPMGR