ACQUITY sample manager fails sample syringe leak test - WKB57007
SYMPTOMS
- Sample syringe leak test failure
- No peaks observed in chromatogram
ENVIRONMENT
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC H-Class
CAUSE
Air in lines
FIX or WORKAROUND
- Tighten the sample syringe by turning it clockwise.
- Reprime the sample syringe 30 times with methanol and verify that the syringe has no air bubbles.
- Repeat the test.
ADDITIONAL INFORMATION
Double-check the seal wash and purge lines and make sure that the filters are fully submerged in solvent. If the filters are above the solvent or air bubbles are seen in the seal wash or purge lines, prime the sample syringe for 30 cycles - or until the air bubbles are purged from the syringe.
See also:
WKB4005 SM-FTN syringe leak test fails intermittently
WKB49303 ACQUITY UPLC sample syringe leak test fails
WKB54861 Sample manager FTN leak test fails
WKB56424 Failing sample syringe leak test on Sample Manager FTN
id57007, A-30SM, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARQSM, UPARSMFTN, UPBINARY, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPPDAARCB, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO