Skip to main content
Waters

ACQUITY sample manager fails sample syringe leak test - WKB57007

Article number: 57007

SYMPTOMS

  • Sample syringe leak test failure
  • No peaks observed in chromatogram

ENVIRONMENT

  • ACQUITY UPLC Sample Manager
  • AQCUITY UPLC M-Class Sample Manager 
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class 

CAUSE

Air in lines

FIX or WORKAROUND

  1. Tighten the sample syringe by turning it clockwise.
  2. Reprime the sample syringe 30 times with methanol and verify that the syringe has no air bubbles.
  3. Repeat the test.
     

ADDITIONAL INFORMATION

Double-check the seal wash and purge lines and make sure that the filters are fully submerged in solvent. If the filters are above the solvent or air bubbles are seen in the seal wash or purge lines, prime the sample syringe for 30 cycles - or until the air bubbles are purged from the syringe.  

See also:

WKB4005 SM-FTN syringe leak test fails intermittently

WKB49303 ACQUITY UPLC sample syringe leak test fails

WKB54861 Sample manager FTN leak test fails

WKB56424 Failing sample syringe leak test on Sample Manager FTN

Not able to find a solution? Click here to request help.